Case study · HPHT reactors & cells

Studying corrosive reactions at 100 bar & 350°C: a custom 2L Hastelloy C276 stirred reactor

For a chemical engineering research laboratory, IRIAN Mecatronics designed and built a pilot test rig to study chemical recycling processes. The equipment features a 2L stirred reactor made of Hastelloy C276, capable of operating up to 100 bar (1450 psi) and 350°C (662°F) while allowing for multiphase injections and sampling during the reaction.

HPHT reactors & cells100 bar350 °C2 LConfidential project · photo withheld

At a glance

Client
Chemical engineering research laboratory
Field
Process development
Technique
Reaction kinetics, Depolymerization, Chemical recycling
Maximum operating pressure100 bar
Design pressure120.6 bar
Operating temperatureAmbient to 350 °C
Reactor volume2 L

01 — The research question

How to instrument and operate a high-pressure reactor to track reaction kinetics in a corrosive environment?

Studying the kinetics of a chemical reaction requires taking representative samples at different stages of the process without disturbing the conditions. This becomes complex when the reaction occurs at high pressure and high temperature, in a corrosive medium, and with potentially viscous fluids. The design must balance material resistance, absolute sealing, operator safety, and measurement reliability. Our client, a university laboratory, needed a versatile tool to develop depolymerization processes. Standard lab reactors could not meet all their constraints: the combination of 100 bar pressure, 350°C temperature, Hastelloy C276 for all wetted parts, agitation for viscous fluids, specific sampling circuits, and safe ergonomics for frequent handling of the vessel. It required a fully custom design, conceived as an integrated system.

Maintaining a perfect seal at 100 bar and 350°C in a stirred reactor, while ensuring simple and safe handling of the heavy vessel for loading, cleaning, and maintenance.
IRIAN Mecatronics · 17 rue du Pont Long, 64160 Morlaàs, France
+33 (0)5 59 90 29 80 · info-meca@irian.fr

02 — Our approach

The design approach

  1. Automating vessel handling for safety and efficiency

    The 2L Hastelloy vessel is too heavy to be handled safely by hand. We designed a motorized system that lowers and pivots the vessel for emptying and cleaning. The lid, which holds all the instrumentation, remains fixed. This approach, chosen over a removable lid, preserves the integrity of the connections and drastically reduces the time between tests.

  2. Designing a modular rig for a constrained installation

    The complete test rig would not fit in the laboratory's service elevator. To avoid exceptional logistics, it was designed in two independent modules (process and control), each respecting the size limits. The two parts were transported separately and reconnected on-site, ensuring a quick and controlled installation.

  3. Ensuring representative sampling without disturbing the process

    To monitor kinetics without disrupting the reaction at 100 bar, we integrated dedicated, isolatable sampling circuits for the liquid and gas phases. A dip tube draws liquid from the heart of the mixture, while a port in the reactor's headspace captures gas. This allows the operator to obtain reliable data in complete safety.

  4. Mastering the thermal cycle to accelerate test campaigns

    The process required heating to 350°C in under an hour and rapid cooling. The solution combines powerful electric heating with high-performance insulation and an external active cooling system. Unlike an internal coil, this solution does not complicate cleaning and allows experiments to be run in quicker succession.

03 — Technical specifications

Maximum operating pressure100 bar (1450 psi)
Design pressure120.6 bar
Operating temperatureAmbient to 350 °C (662 °F)
Reactor volume2 L
Wetted materialsHastelloy C276
Agitation typeMagnetic drive
Agitation speed5 to 1200 rpm
Maximum agitation torque6 N.m
Minimum filling volume1.3 L
CertificationCompliance with Pressure Equipment Directive (PED) 2014/68/EU

04 — Other applications

Does your project resemble this case study?

Fine chemicals and pharmaceuticals

The challenge is often synthesis under pressure in aggressive media, with precise kinetic monitoring. The principle of a special alloy reactor with a magnetic drive and dedicated sampling systems would be retained, adapting the volume and adding compliance with GMP standards or online sensors.

Energy and geosciences (CO2 storage)

For studying fluid-rock interactions under reservoir conditions, the need for corrosion resistance at high pressure and temperature is similar. The overall design would remain, but the reactor geometry could become a coring cell, and the instrumentation could include permeability or acoustic measurements.

Energy and hydrogen (materials, catalysis)

To test materials or catalysts for hydrogen, the rig's general architecture (gas management, P/T control, supervision) is transposable. The changes would concern the materials, to resist hydrogen embrittlement, and the addition of safety features specific to flammable gases (detectors, ventilation, ATEX certification).

A similar test to run?
An engineer from our design office replies as soon as possible.
irian-mecatronics.com